Tonometer readings changed by less than 2 mmHg despite technique variations.

Sensitivity of the iFalcon TM V700 rebound tonometer to measurement distance, angle and position in ex-vivo rabbit eyes.

Jing Jin … Ji Fu
BMC veterinary research · 2026
Background

Rebound tonometers are widely used for intraocular pressure (IOP) monitoring in animal experiments.

Purpose

To evaluate the measurement stability of the iFalcon TM V700 rebound tonometer under different measuring conditions, this study employed a stepwise pressure-increasing protocol to set eight target IOP levels (10-50 mmHg) in two enucleated rabbit eyes (n = 2).

Results

device readings barely shifted from operator handling differences - a good sign

Distance
-1.8125mmHg
Angle
-0.945mmHg
Position
-1.8175mmHg
More results

We systematically investigated the effects of measuring distance (4-7 mm), measuring angle (approximately 30° superior, inferior, left, and right relative to the corneal normal), and corneal measuring location (central cornea and 3 mm peripheral cornea) on IOP readings.

More results

Paired t-tests were used to compare measurements between each test condition and its corresponding reference condition.

“
Conclusion

Although the iFalcon TM V700 exhibited statistically significant dependent biases under the aforementioned operating conditions, the absolute magnitudes were limited and had little impact on experimental interpretation and conclusions, suggesting that the device demonstrates reasonable measurement stability under the tested conditions.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
AI / Informatics
0·962 for OCT
MerMED-FM was very accurate at diagnosing diseases using eye scans
AI / Informatics
0.98
Artificial intelligence's eyelid-height measurements matched doctors' manual measurements almost perfectly
AI / Informatics
92.1%
combining eye scans and photos correctly told benign from cancerous lesions apart nearly every time
Cohort Study
28.6%
recurred locally in more than 1 in 4 patients over years of follow-up
Cohort Study
-0.395
excision group's post-op eyelid fullness score was lower - greater improvement
Cohort Study
86.7%
of infants probed after 12 months still had unresolved tear duct blockage
Observational
16%
eyelid tissue in rosacea patients showed less of this key repair-signaling protein inside cell nuclei
Cohort Study
25%
about 1 in 4 treated cases had symptoms return after improving