Sensitivity of the iFalcon TM V700 rebound tonometer to measurement distance, angle and position in ex-vivo rabbit eyes.

Jing Jin, Hao Cheng, Yunhao Su, Ji Fu
Read paper
OphthalmologyAnimal / preclinical

Tonometer readings changed by less than 2 mmHg despite technique variations.

Sensitivity of the iFalcon TM V700 rebound tonometer to measurement distance, angle and position in ex-vivo rabbit eyes.

Jing Jin … Ji Fu
BMC veterinary research · 2026
Background

Rebound tonometers are widely used for intraocular pressure (IOP) monitoring in animal experiments.

Purpose

To evaluate the measurement stability of the iFalcon TM V700 rebound tonometer under different measuring conditions, this study employed a stepwise pressure-increasing protocol to set eight target IOP levels (10-50 mmHg) in two enucleated rabbit eyes (n = 2).

Results

device readings barely shifted from operator handling differences - a good sign

Distance
-1.8125mmHg
Angle
-0.945mmHg
Position
-1.8175mmHg
More results

We systematically investigated the effects of measuring distance (4-7 mm), measuring angle (approximately 30° superior, inferior, left, and right relative to the corneal normal), and corneal measuring location (central cornea and 3 mm peripheral cornea) on IOP readings.

More results

Paired t-tests were used to compare measurements between each test condition and its corresponding reference condition.

“
Conclusion

Although the iFalcon TM V700 exhibited statistically significant dependent biases under the aforementioned operating conditions, the absolute magnitudes were limited and had little impact on experimental interpretation and conclusions, suggesting that the device demonstrates reasonable measurement stability under the tested conditions.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
Cross-sectional
44.1%
had too much homocysteine in the blood, a far larger share than in dry AMD or healthy eyes
Cohort Study
92.3%
most laser-treated eyes had a favorable eye-structure outcome
Cohort Study
56%
lost three or more lines on the eye-chart vision test
Cohort Study
50%
older patients were more likely to lose three or more lines of vision
Study
81%
trials where mask air leaks were detected blowing toward the eyes
Cohort Study
13.5%
Central retinal artery peak blood-flow speed was 13.5% lower in diabetic patients.
Study
379 +/- 156 microm
lower thickness in the central retina, checked with an eye scan
Case Report
9 of 10 eyes
eyes had better measured vision after treatment