Repeatability of Epithelial Thickness Mapping in Normal and Keratoconus Eyes Using a Novel Hybrid Scheimpflug-Optical Coherence Tomography Device.

Victor Aristide Augustin, Maximilian Friedrich, Julia Keim … Gerd Uwe Auffarth
Read paper
OphthalmologyCross-sectional

Corneal epithelial thickness measurements are slightly less repeatable in keratoconus than healthy eyes

Repeatability of Epithelial Thickness Mapping in Normal and Keratoconus Eyes Using a Novel Hybrid Scheimpflug-Optical Coherence Tomography Device.

Victor Aristide Augustin … Gerd Uwe Auffarth
Cornea · 2026
Purpose

The aim of this study was to analyze the epithelial thickness measurement repeatability in healthy and keratoconus eyes and define severity-dependent thresholds.

Methods

This prospective, cross-sectional, observational, monocentric study included a total of 217 eyes of 217 subjects divided into a keratoconus group (166 eyes of 166 subjects) and a healthy control group (51 eyes of 51 subjects).

n = 217 eyes
Results

keratoconus eyes showed slightly less consistent repeat corneal-surface-layer thickness readings than healthy eyes

3.3RC
Healthy
3.9RC
Keratoconus
More results

The overall median repeatability with (repeatability coefficients = 3.7) and without tear film (repeatability coefficients = 3.8) did not differ significantly (P > 0.05).

Repeatability worsened significantly in severe keratoconus stages (P < 0.01), whereas crosslinking status had only a marginal effect.

“
Conclusion · 1 of 2

The hybrid tomographic device demonstrated high intrasubject repeatability for epithelial thickness assessment in healthy and keratoconus eyes, supporting its technical feasibility for clinical use.

Conclusion · 2 of 2

Further studies are needed to analyze the benefit of incorporating epithelium into established keratoconus scores using the hybrid system.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
AI / Informatics
0·962 for OCT
MerMED-FM was very accurate at diagnosing diseases using eye scans
AI / Informatics
0.98
Artificial intelligence's eyelid-height measurements matched doctors' manual measurements almost perfectly
AI / Informatics
92.1%
combining eye scans and photos correctly told benign from cancerous lesions apart nearly every time
Cohort Study
28.6%
recurred locally in more than 1 in 4 patients over years of follow-up
Cohort Study
-0.395
excision group's post-op eyelid fullness score was lower - greater improvement
Cohort Study
86.7%
of infants probed after 12 months still had unresolved tear duct blockage
Observational
16%
eyelid tissue in rosacea patients showed less of this key repair-signaling protein inside cell nuclei
Cohort Study
25%
about 1 in 4 treated cases had symptoms return after improving