Repeatability of Epithelial Thickness Mapping in Normal and Keratoconus Eyes Using a Novel Hybrid Scheimpflug-Optical Coherence Tomography Device.

Victor Aristide Augustin, Maximilian Friedrich, Julia Keim … Gerd Uwe Auffarth
Read paper
OphthalmologyCross-sectional

Corneal epithelial thickness measurements are slightly less repeatable in keratoconus than healthy eyes

Repeatability of Epithelial Thickness Mapping in Normal and Keratoconus Eyes Using a Novel Hybrid Scheimpflug-Optical Coherence Tomography Device.

Victor Aristide Augustin … Gerd Uwe Auffarth
Cornea · 2026
Purpose

The aim of this study was to analyze the epithelial thickness measurement repeatability in healthy and keratoconus eyes and define severity-dependent thresholds.

Methods

This prospective, cross-sectional, observational, monocentric study included a total of 217 eyes of 217 subjects divided into a keratoconus group (166 eyes of 166 subjects) and a healthy control group (51 eyes of 51 subjects).

n = 217 eyes
Results

keratoconus eyes showed slightly less consistent repeat corneal-surface-layer thickness readings than healthy eyes

3.3RC
Healthy
3.9RC
Keratoconus
More results

The overall median repeatability with (repeatability coefficients = 3.7) and without tear film (repeatability coefficients = 3.8) did not differ significantly (P > 0.05).

Repeatability worsened significantly in severe keratoconus stages (P < 0.01), whereas crosslinking status had only a marginal effect.

“
Conclusion · 1 of 2

The hybrid tomographic device demonstrated high intrasubject repeatability for epithelial thickness assessment in healthy and keratoconus eyes, supporting its technical feasibility for clinical use.

Conclusion · 2 of 2

Further studies are needed to analyze the benefit of incorporating epithelium into established keratoconus scores using the hybrid system.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
Cross-sectional
44.1%
had too much homocysteine in the blood, a far larger share than in dry AMD or healthy eyes
Cohort Study
92.3%
most laser-treated eyes had a favorable eye-structure outcome
Cohort Study
56%
lost three or more lines on the eye-chart vision test
Cohort Study
50%
older patients were more likely to lose three or more lines of vision
Study
81%
trials where mask air leaks were detected blowing toward the eyes
Cohort Study
13.5%
Central retinal artery peak blood-flow speed was 13.5% lower in diabetic patients.
Study
379 +/- 156 microm
lower thickness in the central retina, checked with an eye scan
Case Report
9 of 10 eyes
eyes had better measured vision after treatment