Making Corneal Sensation Testing Accessible: Bench Validation of a Low-Cost Adjustable Monofilament Aesthesiometer.

David Maskill
Read paper
OphthalmologyStudy

Low-cost corneal sensation device costs about 1% of the standard device's price.

Making Corneal Sensation Testing Accessible: Bench Validation of a Low-Cost Adjustable Monofilament Aesthesiometer.

David Maskill
Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
Purpose

This study evaluated whether a generic adjustable monofilament corneal aesthesiometer (generic adjustable monofilament corneal aesthesiometer) fabricated from commodity materials could replicate Cochet-Bonnet performance.

Methods

The generic adjustable monofilament corneal aesthesiometer comprised a 0.12 mm nylon monofilament of adjustable length on a millimetre-scaled sliding carrier, constructed for £10.28.

Results
1%
device works about as cheaply as physical materials alone can allow
More results

Both devices produced monotonically increasing pressure-length relationships consistent with beam mechanics.

Coefficient of variation ranged from 11.0% to 19.9%.

Geometric mean bias was ×1.22 (95% limits of agreement ×0.939 to ×1.576) at 60% humidity and ×1.41 (×1.177 to ×1.680) at 50% humidity.

“
Conclusion · 1 of 2

A corneal aesthesiometer replicating Cochet-Bonnet mechanical principles can be constructed for approximately 1% of its cost, with preserved rank-order fidelity and acceptable repeatability across clinically relevant pressures.

Conclusion · 2 of 2

This device could be suitable for community screening, ward-based settings and resource-scarce environments. Prospective clinical validation of reproducibility and threshold detection in normal and hypoaesthetic corneas are now warranted.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
AI / Informatics
0·962 for OCT
MerMED-FM was very accurate at diagnosing diseases using eye scans
AI / Informatics
0.98
Artificial intelligence's eyelid-height measurements matched doctors' manual measurements almost perfectly
AI / Informatics
92.1%
combining eye scans and photos correctly told benign from cancerous lesions apart nearly every time
Cohort Study
28.6%
recurred locally in more than 1 in 4 patients over years of follow-up
Cohort Study
-0.395
excision group's post-op eyelid fullness score was lower - greater improvement
Cohort Study
86.7%
of infants probed after 12 months still had unresolved tear duct blockage
Observational
16%
eyelid tissue in rosacea patients showed less of this key repair-signaling protein inside cell nuclei
Cohort Study
25%
about 1 in 4 treated cases had symptoms return after improving