Making Corneal Sensation Testing Accessible: Bench Validation of a Low-Cost Adjustable Monofilament Aesthesiometer.

David Maskill
Read paper
OphthalmologyStudy

Low-cost corneal sensation device costs about 1% of the standard device's price.

Making Corneal Sensation Testing Accessible: Bench Validation of a Low-Cost Adjustable Monofilament Aesthesiometer.

David Maskill
Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
Purpose

This study evaluated whether a generic adjustable monofilament corneal aesthesiometer (generic adjustable monofilament corneal aesthesiometer) fabricated from commodity materials could replicate Cochet-Bonnet performance.

Methods

The generic adjustable monofilament corneal aesthesiometer comprised a 0.12 mm nylon monofilament of adjustable length on a millimetre-scaled sliding carrier, constructed for £10.28.

Results
1%
device works about as cheaply as physical materials alone can allow
More results

Both devices produced monotonically increasing pressure-length relationships consistent with beam mechanics.

Coefficient of variation ranged from 11.0% to 19.9%.

Geometric mean bias was ×1.22 (95% limits of agreement ×0.939 to ×1.576) at 60% humidity and ×1.41 (×1.177 to ×1.680) at 50% humidity.

“
Conclusion · 1 of 2

A corneal aesthesiometer replicating Cochet-Bonnet mechanical principles can be constructed for approximately 1% of its cost, with preserved rank-order fidelity and acceptable repeatability across clinically relevant pressures.

Conclusion · 2 of 2

This device could be suitable for community screening, ward-based settings and resource-scarce environments. Prospective clinical validation of reproducibility and threshold detection in normal and hypoaesthetic corneas are now warranted.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
Cross-sectional
44.1%
had too much homocysteine in the blood, a far larger share than in dry AMD or healthy eyes
Cohort Study
92.3%
most laser-treated eyes had a favorable eye-structure outcome
Cohort Study
56%
lost three or more lines on the eye-chart vision test
Cohort Study
50%
older patients were more likely to lose three or more lines of vision
Study
81%
trials where mask air leaks were detected blowing toward the eyes
Cohort Study
13.5%
Central retinal artery peak blood-flow speed was 13.5% lower in diabetic patients.
Study
379 +/- 156 microm
lower thickness in the central retina, checked with an eye scan
Case Report
9 of 10 eyes
eyes had better measured vision after treatment