Repeatability and reproducibility of a clinical device for Brillouin microscopy to measure the biomechanics of the anterior segment of the eye: In vivo tests.

Hannah Pearl Glucksman, Andrew Shin, Lawrence Yoo … Claire Shelley Barnes
Read paper
OphthalmologyCohort study

The Brillouin Optical Scanning System shows low variation in lens stiffness across devices.

Repeatability and reproducibility of a clinical device for Brillouin microscopy to measure the biomechanics of the anterior segment of the eye: In vivo tests.

Hannah Pearl Glucksman … Claire Shelley Barnes
PloS one · 2026
Purpose

We performed a prospective study to assess the repeatability and reproducibility of the novel Brillouin Optical Scanning System (BOSS®) in measuring the biomechanical properties of the cornea and crystalline lens in vivo, to determine device utility in a clinical setting.

Methods

Adult study participants underwent standard ophthalmic exams to confirm eligibility.

n = 33 participants
2.1%
Results
2.1%
this is the low variation in eye lens stiffness when measured by different devices
n = 33 participants
More results

Eleven of the 33 participants (ages 24-72 years; 15 females) had keratoconus (5 with prior crosslinking, 3 without crosslinking, and 3 post-keratoplasty); the other 22 participants were visually normal.

More results

The mean corneal stiffness was 2.83 ± 0.031 GPa, with an overall repeatability coefficients of variation of 1.1%, ranging from 2.85 ± 0.015 with a coefficients of variation of 0.5% to 2.81 ± 0.036 with a coefficients of variation of 1.3% across devices.

“
Conclusion · 1 of 2

The study results demonstrated consistent stiffness values for both lenses and corneas within and between the three BOSS devices tested.

Conclusion · 2 of 2

The study shows that this device could be a useful new biomechanics-measuring tool for applications in ophthalmology.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
AI / Informatics
0·962 for OCT
MerMED-FM was very accurate at diagnosing diseases using eye scans
AI / Informatics
0.98
Artificial intelligence's eyelid-height measurements matched doctors' manual measurements almost perfectly
AI / Informatics
92.1%
combining eye scans and photos correctly told benign from cancerous lesions apart nearly every time
Cohort Study
28.6%
recurred locally in more than 1 in 4 patients over years of follow-up
Cohort Study
-0.395
excision group's post-op eyelid fullness score was lower - greater improvement
Cohort Study
86.7%
of infants probed after 12 months still had unresolved tear duct blockage
Observational
16%
eyelid tissue in rosacea patients showed less of this key repair-signaling protein inside cell nuclei
Cohort Study
25%
about 1 in 4 treated cases had symptoms return after improving