Repeatability and reproducibility of a clinical device for Brillouin microscopy to measure the biomechanics of the anterior segment of the eye: In vivo tests.

Hannah Pearl Glucksman, Andrew Shin, Lawrence Yoo … Claire Shelley Barnes
Read paper
OphthalmologyCohort study

The Brillouin Optical Scanning System shows low variation in lens stiffness across devices.

Repeatability and reproducibility of a clinical device for Brillouin microscopy to measure the biomechanics of the anterior segment of the eye: In vivo tests.

Hannah Pearl Glucksman … Claire Shelley Barnes
PloS one · 2026
Purpose

We performed a prospective study to assess the repeatability and reproducibility of the novel Brillouin Optical Scanning System (BOSS®) in measuring the biomechanical properties of the cornea and crystalline lens in vivo, to determine device utility in a clinical setting.

Methods

Adult study participants underwent standard ophthalmic exams to confirm eligibility.

n = 33 participants
2.1%
Results
2.1%
this is the low variation in eye lens stiffness when measured by different devices
n = 33 participants
More results

Eleven of the 33 participants (ages 24-72 years; 15 females) had keratoconus (5 with prior crosslinking, 3 without crosslinking, and 3 post-keratoplasty); the other 22 participants were visually normal.

More results

The mean corneal stiffness was 2.83 ± 0.031 GPa, with an overall repeatability coefficients of variation of 1.1%, ranging from 2.85 ± 0.015 with a coefficients of variation of 0.5% to 2.81 ± 0.036 with a coefficients of variation of 1.3% across devices.

“
Conclusion · 1 of 2

The study results demonstrated consistent stiffness values for both lenses and corneas within and between the three BOSS devices tested.

Conclusion · 2 of 2

The study shows that this device could be a useful new biomechanics-measuring tool for applications in ophthalmology.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
Cross-sectional
44.1%
had too much homocysteine in the blood, a far larger share than in dry AMD or healthy eyes
Cohort Study
92.3%
most laser-treated eyes had a favorable eye-structure outcome
Cohort Study
56%
lost three or more lines on the eye-chart vision test
Cohort Study
50%
older patients were more likely to lose three or more lines of vision
Study
81%
trials where mask air leaks were detected blowing toward the eyes
Cohort Study
13.5%
Central retinal artery peak blood-flow speed was 13.5% lower in diabetic patients.
Study
379 +/- 156 microm
lower thickness in the central retina, checked with an eye scan
Case Report
9 of 10 eyes
eyes had better measured vision after treatment