Plaque-adaptive flexible film dosimetry system for Ru-106 plaque brachytherapy commissioning and quality assurance.

Jin Dong Cho, Hojae Kim, Dong Wook Kim … Jung-In Kim
Read paper
OphthalmologyStudy

New flexible film dosimeter matched reference doses within 3% for Ru-106 plaques.

Plaque-adaptive flexible film dosimetry system for Ru-106 plaque brachytherapy commissioning and quality assurance.

Jin Dong Cho … Jung-In Kim
Medical physics · 2026
Background

Ruthenium-106 (Ru-106) plaque brachytherapy is an established treatment modality for choroidal melanoma that delivers localized beta radiation while sparing surrounding ocular structures.

Purpose

This study aimed to assess the feasibility and dosimetric accuracy of a plaque-adaptive flexible film dosimeter (PFD) for planar and central axis (central axis) dose measurements in Ru-106 plaque brachytherapy.

Methods

The PFD consisted of an active layer of lithium salt of pentacosa-10,12-diynoic acid (LiPCDA) enclosed between upper and lower silicone sheets.

Results

measured doses landed within about 3% of true values on average

CCA
2.1%
COB
2.1%
CIB
2.2%
More results

In the high-dose central region, the mean differences for CCA, COB, and CIB were 2.1% ± 0.8%, 2.1% ± 0.8%, and 2.2% ± 0.7%, respectively.

Central axis dose profiles measured with a microDiamond detector differed by 0.89% ± 0.58%.

Total dose uncertainty ranged from 4.43% at 0.3 Gy to 4.19% at 2.4 Gy.

“
Conclusion · 1 of 3

The PFD enabled accurate and high-resolution dose measurements on the curved surfaces of Ru-106 plaques.

Conclusion · 2 of 3

Both planar and central axis measurements demonstrated strong agreement with manufacturer-certified reference data across all plaque types.

Conclusion · 3 of 3

This system offers a practical and reproducible solution for commissioning and quality assurance in Ru-106 plaque brachytherapy, supporting its clinical integration for choroidal melanoma treatment.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
Cross-sectional
44.1%
had too much homocysteine in the blood, a far larger share than in dry AMD or healthy eyes
Cohort Study
92.3%
most laser-treated eyes had a favorable eye-structure outcome
Cohort Study
56%
lost three or more lines on the eye-chart vision test
Cohort Study
50%
older patients were more likely to lose three or more lines of vision
Study
81%
trials where mask air leaks were detected blowing toward the eyes
Cohort Study
13.5%
Central retinal artery peak blood-flow speed was 13.5% lower in diabetic patients.
Study
379 +/- 156 microm
lower thickness in the central retina, checked with an eye scan
Case Report
9 of 10 eyes
eyes had better measured vision after treatment