Dosimetric feasibility of carbon-ion radiotherapy for ocular melanomas with local effect model-based treatment planning system.

Yuze Yang, Jiayao Sun, Xue Ming … Yinxiangzi Sheng
Read paper
DermatologyStudy

A conversion factor of 0.65 was needed to adapt carbon-ion radiotherapy dose planning for eye melanoma.

Dosimetric feasibility of carbon-ion radiotherapy for ocular melanomas with local effect model-based treatment planning system.

Yuze Yang … Yinxiangzi Sheng
Journal of radiation research · 2026
Purpose

This study evaluates the dosimetric feasibility of carbon-ion radiation therapy (carbon-ion radiation therapy) for ocular melanoma (ocular melanoma) using a local effect model-I-based treatment planning system (treatment planning system) combined with pencil beam scanning delivery.

0.65
Results
0.65
converts standard dose planning to match this therapy - lower means a smaller adjusted dose
More results

The 3D gamma passing rate between treatment planning system and MC was ≥94% using 3%-3 mm criteria.

Mean dose deviations between the treatment planning system and measurement were ≤±1.61% for all targets sizes except the smallest one (7 mm).

More results

Mean distance-to-agreement was <1 mm for all plans except one with a range shifter (distance-to-agreement = 2.92 mm).

Phantom cases showed a lateral penumbra of 5.4-6.2 mm, and distal dose fall-off measures 3.6 and 5.6 mm.

“
Conclusion

The results underscore the need for relative biological effectiveness-weighted dose conversion when applying Quantum Science and Technology's carbon-ion radiation therapy experience to ocular melanoma treatment.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
Cross-sectional
44.1%
had too much homocysteine in the blood, a far larger share than in dry AMD or healthy eyes
Cohort Study
92.3%
most laser-treated eyes had a favorable eye-structure outcome
Cohort Study
56%
lost three or more lines on the eye-chart vision test
Cohort Study
50%
older patients were more likely to lose three or more lines of vision
Study
81%
trials where mask air leaks were detected blowing toward the eyes
Cohort Study
13.5%
Central retinal artery peak blood-flow speed was 13.5% lower in diabetic patients.
Study
379 +/- 156 microm
lower thickness in the central retina, checked with an eye scan
Case Report
9 of 10 eyes
eyes had better measured vision after treatment