Randomized trialJournal of glaucoma · 2011

Goldmann applanation tonometry and dynamic contour tonometry are not correlated with central corneal thickness in primary open angle glaucoma.

Guy Regev, Alon Harris, Brent Siesky … Rita Ehrlich
Read paper
OphthalmologyRandomized trial

Dynamic contour tonometry readings were not correlated with central corneal thickness.

Goldmann applanation tonometry and dynamic contour tonometry are not correlated with central corneal thickness in primary open angle glaucoma.

Guy Regev … Rita Ehrlich
Journal of glaucoma · 2011
Purpose

To compare intraocular pressure (IOP) measured by Goldmann applanation tonometry (goldmann applanation tonometry) and dynamic contour tonometry (dynamic contour tonometry) and assess their relationship to central corneal thickness (central corneal thickness) in patients with primary open angle glaucoma (open angle glaucoma).

Methods

Goldmann applanation tonometry and dynamic contour tonometry were measured in a randomized order followed by central corneal thickness (ultrasonic corneal pachymetry) during a single study visit.

n = 116 patients
r=0.05
Results
r=0.05
Corneal thickness was not associated with dynamic contour intraocular pressure readings.
n = 116 patients
More results

IOP values obtained by dynamic contour tonometry and goldmann applanation tonometry showed a strong positive correlation in patients with open angle glaucoma (r=0.93; P<0.001).

Mean IOP measured with dynamic contour tonometry [18.4 (5.1) mm Hg] was significantly higher (P<0.001) than goldmann applanation tonometry IOP measurements [16.5 (4.5) mmHg].

More results

The Bland-Altman data showed that the amount of disagreement between IOP assessment techniques varied, suggesting a proportional bias.

“
Conclusion · 1 of 4

In this group of patients with open angle glaucoma, there was a strong correlation between goldmann applanation tonometry and dynamic contour tonometry measurements of IOP.

Conclusion · 2 of 4

IOP measured with dynamic contour tonometry was consistently higher than IOP measured with goldmann applanation tonometry.

Conclusion · 3 of 4

Neither goldmann applanation tonometry nor dynamic contour tonometry measurements were correlated with central corneal thickness.

Conclusion · 4 of 4

This data suggests that factors other than central corneal thickness may be involved in the tendency of dynamic contour tonometry to produce higher measures of IOP than goldmann applanation tonometry.

Read paper
0 comments

No comments yet. Be the first.

Related papers

LatestFoundational
Cross-sectional
44.1%
had too much homocysteine in the blood, a far larger share than in dry AMD or healthy eyes
Cohort Study
92.3%
most laser-treated eyes had a favorable eye-structure outcome
Cohort Study
56%
lost three or more lines on the eye-chart vision test
Cohort Study
50%
older patients were more likely to lose three or more lines of vision
Study
81%
trials where mask air leaks were detected blowing toward the eyes
Cohort Study
13.5%
Central retinal artery peak blood-flow speed was 13.5% lower in diabetic patients.
Study
379 +/- 156 microm
lower thickness in the central retina, checked with an eye scan
Case Report
9 of 10 eyes
eyes had better measured vision after treatment